| Automatic Buffer Recognition: |
Yes |
| Backlight: |
Yes |
| Barometric Pressure Measurement: |
Automatic compensation of DO when using an LDO or LBOD probe |
| BOD5/CBOD resolution: |
Available when used with Hach WIMS BOD Manager software |
| BOD Bottle: |
300 mL/100 pcs |
| Calibration: |
Demal (1D/ 0.1D/ 0.01D);
Molar (0.1M/ 0.01M/0.001M);
NaCl (0.05%; 25µS/cm; 1000µS/cm; 18mS/cm);
Standard sea water;
User defined |
| Calibration Intervals/Alerts/Reminder: |
Off, selectable from 2 hours to 7 days |
| Communication: |
Integrated USB type A (for USB 2.0 flash memory device, printer, keyboard) and Integrated USB type B (for PC) |
| Compliance Certifications: |
CE.WEEE |
| Conductivity measurement: |
Yes |
| Conductivity measurement: Temperature correction: |
None; Linear; NaCl Non-Linear Natural Water. |
| Conductivity Measurement Range: |
0.01 µS/cm - 200.0 mS/cm |
| Conductivity resolution: |
0.01 µS/cm - 0.1 mS/cm upon selected measuring range |
| Contents: |
Meter + Probe(s) + Lab. Accessories |
| Custom Calibration Standards: |
Yes |
| Data Export: |
Download via USB connection to PC or flash memory device. Automatically transfer entire data log or as readings are taken |
| Data Memory: |
500 records/FIFO |
| Data Storage: |
Automatic, GLP ISO compliant reading data stored with calibration details. |
| Dimensions: |
859 x 175 x 235 mm (3.38 x 6.88 x 9.25 in.) |
| Display: |
Detailed mode/Large mode |
| Display Type: |
440 x 160 pixel LCD with backlight illumination |
| DO Measurement Range: |
0.1 - 20.0 mg/L (ppm) 1 - 200% saturation |
| DO Resolution: |
0.1 |
| DO sensor calibration: |
• 100% (water-saturated air (100%) calibration
• 100% with 0 (water-saturated air (100%) calibration with 0 point
• mg/L (calibration with a specified dissolved oxygen concentration (mg/L) solution)
• mg/L with 0 (calibration with a specified dissolved oxygen concentration (mg/L) solution with 0 point)
• Factory (calibration with the default LDO calibration) |
| Electrode Holder: |
NA |
| Electrode stand: |
Optional |
| GLP Features: |
Date; Time; Sample ID; Operator ID |
| Instrument: |
Bench |
| IP Rating: |
IP54 (Resistant to spray of water; Dust-proof) |
| ISE Direct Measurement Range: |
Yes |
| ISE Electrode Calibration: |
2 - 5 points |
| Kit?: |
Yes |
| Languages: |
English, French, German, Italian, Spanish, Danish, Dutch, Polish, Portuguese, Turkish, Swedish, Czech, Russian |
| Lock Function: |
Continuous / Auto-stabilization ("press to read") / At Interval |
| Measurement Method: |
Luminescent Dissolved Oxygen |
| mV Measurement at Stable Reading: |
Yes |
| mV Measurement Range: |
-1500 - 1500 mV |
| mV Resolution: |
0.1 mV |
| Needed Cable Length: |
1 |
| Operating Error Messages: |
Clear text error messages displayed |
| Operating Humidity: |
90 % relative humidity (non-condensing) |
| Operating Interface: |
Soft Touch Keypad |
| Operating Temperature: |
0 - +60 °C (32 - 140 °F) |
| ORP Electrode Calibration: |
Predefined ORP standards (including Zobell's solution) |
| Parameter: |
pH/Oxydo Reduction Potential (ORP)
Conductivity/Total Dissolved Solid (TDS)/Salinity/Resistivity
Dissolved Oxygen (DO)
Biochemical Oxygen Demand (BOD)
Ion Selective Electrode (ISE): Ammonia, Ammonium, Chloride, Fluoride, Nitrate, Sodium |
| PC Data Transfer Software: |
HQd Series Meter Data Transfer Utility |
| pH Buffer Sets: |
Color-coded: 4.01, 7.00, 10.01 pH;
IUPAC: 1.679, 4.005, 7.000, 10.012, 12.45 pH
DIN: 1.09, 4.65, 9.23 pH
User-defined custom buffer sets |
| pH Electrode calibration: |
1 - 3 Calibration points
Calibration summary data logged and displayed |
| pH Measurement Range: |
0 - 14 pH |
| pH Resolution: |
Selectable:
0.001/0.01/0.1 pH |
| Printer: |
Yes, Optional |
| Probes included: |
LBOD10101 |
| Range: |
0.05 - 20.00 mg/L |
| Region: |
US |
| Resistivity Measurement: |
2.5 Ω.cm - 49 MΩ.cm |
| Resistivity Resolution: |
0.1 |
| Salinity Measurement Range: |
0 - 42 (ppt) (‰) |
| Salinity Resolution: |
0.01 (ppt) (‰) |
| Sensor A: |
LBOD10101 |
| Sensor B: |
NA
|
| Sensor C: |
NA
|
| Sensors: |
Sensor A: LBOD10101
Sensor B: NA
Sensor C: NA |
| TDS Measurement Range: |
0.00 mg/L - 50.0 g/L as NaCl |
| TDS Resolution: |
0.01 mg/L - 0.1 g/L upon measuring range. |
| Temperature Compensation: |
Automatic Temperature compensation for pH |
| Temperature Measurement: |
°C or °F |
| Temperature Range: |
-10°C - 110 °C |
| Temperature Resolution: |
0.1 |
| Type of measurement: |
Lab Wasterwater Monitoring |
| Weight: |
750 g (1.65 lb) without batterie; 850 g (1.87 lb) with batteries |